Electrodes, Sensors, Accessories \ Industrial Electrodes \ pH Glass Reference Electrodes
SE 20 I
  SE 20, SE 21, SE 23
Application Reference electrodes with internal Ag/AgCl reference system are used inconnection with single pH, redox or ion-selective measurement electrodes. SE 23 features low maintenance, application without external pressure compensation and 3 ceramic diaphragms for minimal flow dependency.
Specification -5 ... 80 °C
glass shaft Ø 12 mm
immersion length 120 mm

SE 20:
refillable electrolyte sat. KCl, detachable sleeve junction, glass shaft with side-arm

SE 21:
refillable electrolyte sat. KCl, ceramic diaphragm, glass shaft with side-arm

SE 23:
gel electrolyte (non-refillable), 3 ceramic diaphragms, max. 6 bar

SE 20 K, SE 21 K, SE 23 K:
fixed cable 5 m, without connector

SE 23 I: PG 13.5 threaded plug head S7-system
  Cable Connections and Connectors
GA 151 I
  GA 151
Application This glass-bodied screened pH sensing half cell must be used in conjunction with a reference elcectrode for general precision pH measurement. Special version with extra thick pH glass membrane features application in abrasive samples.
Specification -5 ... 80 °C
pH 0 ... 14
max. 6 bar
isopotential point pH 7 Ag
glass shaft Ø 12 mm
immersion length 120 mm
cylinder membrane

GA 151 K: fixed cable 5 m without connector

GA 151 I: PG 13.5 threaded plug head S7-system
  Anschlussvarianten und Steckverbindungen