LM3000

  • Measurement of pH/ISE/mV, conductivity, chlorine and dissolved oxygen
  • Up to 4 measurement modules (inputs) in a variety of configurations
  • PC software with extensive operating functions
  • Coupling of autosampler and burettes for dosing and automated measurements
  • Virtual channels to calculate different parameters from the measured value
  • Logbook, i. e. documentation of setting changes
  • Plain text menu

Features
  • 5.7” RGB TFT display with touch control
  • Special electrode test for pH
  • Current outputs for each parameter
  • Additional modules for current output possible
  • Timer function
  • Alarm/limit value function
  • 2 PID regulators
  • Data storage and data recording
  • Data transfer with RS232/USB or Ethernet
  • Access control by password

Description

The multiparameter laboratory measuring instrument LM3000 is the complete solution for the tasks of measurement, control and regulation in the laboratory and technical center. Up to 4 internal measurement modules can determine any combination of pH, redox potential, ISE, conductivity, dissolved oxygen, chlorine, temperature and other parameters. This is based on a modular system structure that links the device components via a bus connection.


Specifications
Measuring parameters
  • pH / redox potential (ORP)
  • conductivity
  • dissolved oxygen concentration
  • ion concentration
  • temperature and more

Temperature compensation
  • automatically with Pt 1000
PC interface
  • USB, Ethernet

Power supply
  • 12 ... 24 V AC/DC plug-in mains unit

Ambient temperature
  • -10 ... 50 °C
Enclosure
  • aluminum console housing
  • protection IP 41

Dimension
  • 345x73x170 mm (WxHxD )

Sensor connection
  • 2 x 4 mm Banana plug / BNC plug

Overview system componnents

Actor Modules (internal) Description

DAC3000LM 4 Current outputs
  • 0(4)...20 mA
  • free scaleable

Measuring Modules (internal) Measuring range

MVM3210LM pH value or redoxpotential or concentration (ISE) and temperature
  • -2.000 ... 16.000
  • -2000 ... 2000 mV
  • -10.0 ... 130.0 °C
MVM3220LM Conductivity (2 electrode cells) and temperature
  • 0 ... 100.0 mS/cm auto range
  • -10.0 ... 130.0 °C
MVM3225LM Conductivity (4 electrode cells) and temperature
  • 0 ... 500.0 mS/cm auto range
  • -10.0 ... 130.0 °C
MVM3230LM Dissolved Oxygen and temperature
  • 0 ... 20.00 mg/l
  • 0 ... 200.0 % saturation
  • -10.0 ... 130.0 °C
MVM3260LM Chlorine and temperature
  • 0(4) ... 20 mA
  • free programmable slope / concentration
  • -10.0 ... 130.0 °C
MVM3260ALM "linear characteristic" for Sensors with voltage output
  • 0 ... 5 V DC
  • free programmable
  • -10.0 ... 130.0 °C
MVM3260BLM "linear characteristic" for Sensors with active current output
  • 0(4) ... 20 mA
  • free programmable
  • -10.0 ... 130.0 °C
MVM3260CLM "linear characteristic" for Sensors with passive current output
  • 0(4) ... 20 mA
  • free programmable
  • 15 VDC / 65 mA power supply
  • -10.0 ... 130.0 °C
MVM3270LM Flow for flow sensor with frequency output
  • 0 ... 3 kHz
  • free programmable slope
  • -10.0 ... 130.0 °C