Sensortechnik Meinsberg - Measuring Transmitters
Contact Home Sitemap Search
DE
Products

Measuring Transmitter MV 3000 / 4000

Downloads:
  | Catalog MV 3000/4000-Series |
  | Data Sheets Transmitters |
  | Manual MV 3000/4000-Series |
  | Software DinModule |
Network Measuring Transmitter with AZ 3000

Applications
  • Water Technology
  • Water Quality Monitoring Systems
  • Process Monitoring
  • Process Control

Features
  • cost-efficient measurement of process parameters
  • display for measuring values and clear calibration instructions
  • simultaneous temperature measurement
  • 2 free scaleable analogue outputs and 1 relay output
  • USB interface and PC software
  • modular DIN rail mounted case
  • networking with display unit for 8 measuring values
  • firmware update via USB

Description

The transmitter series MV 3000 and MV 4000 are excellently suited for continuous measurements of one or more parameters in connection with process control units, data loggers, -printers or visualisation systems. They are characterised by high flexibility and simple handling. Calibrations can be done directly on transmitter. An additional display on MV 4000 simplifies measuring value checks, calibrations and makes each transmitter to a stand-alone measuring instrument.
All transmitters can be connected with the display unit AZ 3000. In this way the customer gets a complete multiparameter measuring system with data logger and fieldbus interface RS-485.
Simple software features configuring with logging of configuration parameters, calibration as well as displaying and recording of measuring values directly on PC using USB interface. Additionally the series MV 4000 allows actualisation of firmware via flash update.


Specifications - MV 3000 / 4000
Temperature compensation
  • automatically with Pt 1000 or fixed temperature

Kalibrierung
  • using 2 configurable buttons or PC software

Signalisation
  • LED: measuring value stability for calibration, measuring, errors
Display (Series MV 4000)
  • graphic LCD, 48 x 64 pixel, back-lighted,
  • display for 3 values (value 1, value 2 and temperature value)
  • clear calibration instructions

Controller outputs
  • 1 limit relay, NOC, 24 V AC/DC, max. 1 A
Analogue outputs
  • 2 x 0(4) ... 20 mA oder 2 x 0 ... 5 V

Interfaces
  • MV 4000: USB (optional RS-232)
  • MV 3000: RS-232
  • for configuration, calibration and data logger

Dimensions
  • 22.5x99x114.5 mm (W x H x D)

Specifications - General
Firmware update
  • via PC-interface (only MV 4000)

Power supply
  • 15 ... 24 V AC / DC, ca. 1.5 VA
Enclosure
  • plastic case for mounting on rail tracks DIN EN 50022-35
  • protection IP 40
EMC
  • EN 61326 class B

Overview system componnents - MV 3000 and MV 4000

Type Component Description

MV 40XX Transmitter
  • display
  • USB interface
  • scaleable current outputs
  • 2 x 0(4) ... 20 mA
  • 1 floating relay output
MV 40XX-U Transmitter
  • display
  • USB interface
  • scaleable voltage outputs
  • 2 x 0 ... 5 V
  • 1 floating relay output
MV 30XX Transmitter
  • RS-232 interface
  • scaleable current outputs
  • 2 x 0(4) ... 20 mA
  • 1 floating relay output
MV 30XX-U Transmitter
  • RS-232 interface
  • scaleable voltage outputs
  • 2 x 0 ... 5 V
  • 1 floating relay output

Overview Transmitters

with Display without Display Measuring range Resolution Temperature measurement

MV 4010 MV 3010 pH value
  • pH -2 ... 16
pH 0,01 -10 ... 130 °C
res. 0,1 °C
MV 4015 MV 3015 Redox potential
  • -2000 ... +2000 mV
  -10 ... 130 °C
res. 0,1 °C
MV 4016 MV 3016 Ion concentration
  • sensor specific (ISE)
pH 0,01 -10 ... 130 °C
res. 0,1 °C
MV 4020 MV 3020 Conductivity (2-pol)
  • 0 ... 200 µS/cm
  • 0 ... 2000 µS/cm
  • 0 ... 20 mS/cm
  • 0 ... 100 mS/cm
  • auto range
  • salinity 2 ... 42 g/kg
  • resistor
0,1 µS/cm
1,0 µS
0,01 mS/cm
0,1 mS/cm
-10 ... 130 °C
res. 0,1 °C
MV 4025 MV 3025 Conductivity (4-pol)
  • 0 ... 200 µS/cm
  • 0 ... 2000 µS/cm
  • 0 ... 20 mS/cm
  • 0 ... 500 mS/cm
  • auto range
  • salinity 2 ... 42 g/kg
  • resistor
0,1 µS/cm
1,0 µS
0,01 mS/cm
0,1 mS/cm
-10 ... 130 °C
res. 0,1 °C
MV 4030 MV 3030 Dissolved oxygen
  • 0 ... 200 %
  • 0 ... 20 mg/l
0,1 %
0,01 mg/l
-10 ... 130 °C
res. 0,1 °C
MV 4060 MV 3060 Concentration (chlorine, dioxide, ...)